Convert laboratory instrument output files (PDF, CSV, Excel, TXT) to Allotrope Simple Model (ASM) JSON format or flattened 2D CSV. Use this skill when scientists need to standardize instrument data for LIMS systems, data lakes, or downstream analysis. Supports auto-detection of instrument types. Outputs include full ASM JSON, flattened CSV for easy import, and exportable Python code for data engineers. Common triggers include converting instrument files, standardizing lab data, preparing data for upload to LIMS/ELN systems, or generating parser code for production pipelines.

Use this Skill: https://skilld.dev/gh/anthropics/knowledge-work-plugins/instrument-data-to-allotrope

This session only. Nothing lands on disk.

referencesfield_classification_guide.md

≈4.4k tokens on demand. Your agent reads this file only when SKILL.md points to it.

Field Classification Guide

This guide helps classify instrument data fields into the correct ASM document locations. Use this when mapping raw instrument output to Allotrope Simple Model structure.

ASM Document Hierarchy

<technique>-aggregate-document
├── device-system-document          # Instrument hardware info
├── data-system-document            # Software/conversion info
├── <technique>-document[]          # Per-run/sequence data
│   ├── analyst                     # Who performed the analysis
│   ├── measurement-aggregate-document
│   │   ├── measurement-time
│   │   ├── measurement-document[]  # Individual measurements
│   │   │   ├── sample-document
│   │   │   ├── device-control-aggregate-document
│   │   │   └── [measurement fields]
│   │   └── [aggregate-level metadata]
│   ├── processed-data-aggregate-document
│   │   └── processed-data-document[]
│   │       ├── data-processing-document
│   │       └── [processed results]
│   └── calculated-data-aggregate-document
│       └── calculated-data-document[]

Field Classification Categories

1. Device/Instrument Information → device-system-document

Hardware and firmware details about the physical instrument.

Field Type ASM Field Examples
Instrument name model-number "Vi-CELL BLU", "NanoDrop One"
Serial number equipment-serial-number "VCB-12345", "SN001234"
Manufacturer product-manufacturer "Beckman Coulter", "Thermo Fisher"
Firmware version firmware-version "v2.1.3"
Device ID device-identifier "Instrument_01"
Brand brand-name "Beckman Coulter"

Rule: If the value describes the physical instrument and doesn't change between runs, it goes in device-system-document.


2. Software/Data System Information → data-system-document

Information about software used for acquisition, analysis, or conversion.

Field Type ASM Field Examples
Software name software-name "Chromeleon", "Gen5"
Software version software-version "7.3.2"
File name file-name "experiment_001.xlsx"
File path file-identifier "/data/runs/2024-01-15/"
Database ID ASM-converter-name "allotropy v0.1.55"

Rule: If the value describes software, file metadata, or data provenance, it goes in data-system-document.


3. Sample Information → sample-document

Metadata about the biological/chemical sample being analyzed.

Field Type ASM Field Examples
Sample ID sample-identifier "Sample_A", "LIMS-001234"
Sample name written-name "CHO Cell Culture Day 5"
Sample type/role sample-role-type "unknown sample role", "control sample role"
Batch ID batch-identifier "Batch-2024-001"
Description description "Protein expression sample"
Well position location-identifier "A1", "B3"

Rule: If the value identifies or describes what was measured (not how), it goes in sample-document.


4. Device Control Settings → device-control-aggregate-document

Instrument settings and parameters used during measurement.

Field Type ASM Field Examples
Injection volume sample-volume-setting 10 µL
Wavelength detector-wavelength-setting 254 nm
Temperature compartment-temperature 37°C
Flow rate flow-rate 1.0 mL/min
Exposure time exposure-duration-setting 500 ms
Detector gain detector-gain-setting 1.5
Illumination illumination-setting 80%

Rule: If the value is a configurable instrument parameter that affects measurement, it goes in device-control-aggregate-document.


5. Environmental Conditions → device-control-document or technique-specific

Ambient or controlled environmental parameters during measurement.

Field Type ASM Field Examples
Ambient temperature ambient-temperature 22.5°C
Humidity ambient-relative-humidity 45%
Column temperature compartment-temperature 30°C
Sample temperature sample-temperature 4°C
Electrophoresis temp (technique-specific) 26.4°C

Rule: Environmental conditions that affect measurement quality go with device control or in technique-specific locations.


6. Raw Measurement Data → measurement-document

Direct instrument readings - the "ground truth" data.

Field Type ASM Field Examples
Absorbance absorbance 0.523 AU
Fluorescence fluorescence 12500 RFU
Cell count total-cell-count 2.5e6 cells
Peak area peak-area 1234.5 mAU·min
Retention time retention-time 5.67 min
Ct value cycle-threshold-result 24.5
Concentration (measured) mass-concentration 1.5 mg/mL

Rule: If the value is a direct instrument reading that wasn't computed from other values in this analysis, it goes in measurement-document.


7. Calculated/Derived Data → calculated-data-aggregate-document

Values computed from raw measurements.

Field Type ASM Field Examples
Viability % calculated-result 95.2%
Concentration (from std curve) calculated-result 125 ng/µL
Ratio (260/280) calculated-result 1.89
Relative quantity calculated-result 2.5x
% Recovery calculated-result 98.7%
CV% calculated-result 2.3%

Calculated data document structure:

{
  "calculated-data-name": "viability",
  "calculated-result": {"value": 95.2, "unit": "%"},
  "calculation-description": "viable cells / total cells * 100"
}

Rule: If the value was computed from other measurements in this analysis, it goes in calculated-data-aggregate-document. Include calculation-description when possible.


8. Processed/Analyzed Data → processed-data-aggregate-document

Results from data processing algorithms (peak integration, cell classification, etc.).

Field Type ASM Field Examples
Peak list peak-list Integrated peak results
Cell size distribution cell-diameter-distribution Histogram data
Baseline-corrected data (in processed-data-document) Corrected spectra
Fitted curve (in processed-data-document) Standard curve fit

Associated data-processing-document:

{
  "cell-type-processing-method": "trypan blue exclusion",
  "cell-density-dilution-factor": {"value": 2, "unit": "(unitless)"},
  "minimum-cell-diameter-setting": {"value": 5, "unit": "µm"},
  "maximum-cell-diameter-setting": {"value": 50, "unit": "µm"}
}

Rule: If the value results from an algorithm or processing method applied to raw data, it goes in processed-data-aggregate-document with its processing parameters in data-processing-document.


9. Timing/Timestamps → Various locations

Timestamp Type Location ASM Field
Measurement time measurement-document measurement-time
Run start time analysis-sequence-document analysis-sequence-start-time
Run end time analysis-sequence-document analysis-sequence-end-time
Data export time data-system-document (custom)

Rule: Use ISO 8601 format: 2024-01-15T10:30:00Z


10. Analyst/Operator Information → <technique>-document

Field Type ASM Field Examples
Operator name analyst "jsmith"
Reviewer (custom or extension) "Pending"

Rule: Analyst goes at the technique-document level, not in individual measurements.


Decision Tree

Is this field about...

THE INSTRUMENT ITSELF?
├── Hardware specs → device-system-document
└── Software/files → data-system-document

THE SAMPLE?
└── Sample ID, name, type, batch → sample-document

INSTRUMENT SETTINGS?
└── Configurable parameters → device-control-aggregate-document

ENVIRONMENTAL CONDITIONS?
└── Temp, humidity, etc. → device-control-document

A DIRECT READING?
└── Raw instrument output → measurement-document

A COMPUTED VALUE?
├── From other measurements → calculated-data-document
└── From processing algorithm → processed-data-document

TIMING?
├── When measured → measurement-document.measurement-time
└── When run started/ended → analysis-sequence-document

WHO DID IT?
└── Operator/analyst → <technique>-document.analyst

Common Instrument-to-ASM Mappings

Note: These mappings are derived from the Benchling allotropy library. For authoritative mappings, consult the parser source code for your specific instrument.

Cell Counter (Vi-CELL BLU)

Source: allotropy/parsers/beckman_vi_cell_blu/vi_cell_blu_structure.py

Instrument Field ASM Field
Sample ID sample_identifier
Analysis date/time measurement_time
Analysis by analyst
Viability (%) viability
Viable (x10^6) cells/mL viable_cell_density
Total (x10^6) cells/mL total_cell_density
Cell count total_cell_count
Viable cells viable_cell_count
Average diameter (μm) average_total_cell_diameter
Average viable diameter (μm) average_live_cell_diameter
Average circularity average_total_cell_circularity
Cell type cell_type_processing_method (data-processing)
Dilution cell_density_dilution_factor (data-processing)
Min/Max Diameter minimum/maximum_cell_diameter_setting (data-processing)

Spectrophotometer (NanoDrop)

Instrument Field ASM Field
Sample Name sample_identifier
A260, A280 absorbance (with wavelength)
Concentration mass_concentration
260/280 ratio a260_a280_ratio
Pathlength pathlength

Plate Reader

Instrument Field ASM Field
Well location_identifier
Sample Type sample_role_type
Absorbance/OD absorbance
Fluorescence fluorescence
Plate ID container_identifier

Chromatography (HPLC)

Instrument Field ASM Field
Sample ID sample_identifier
Injection Volume injection_volume
Retention Time retention_time
Peak Area peak_area
Peak Height peak_height
Column Temp column_oven_temperature
Flow Rate flow_rate

Unit Handling

Only use units explicitly present in source data. If a value has no unit specified:

  • Use (unitless) as the unit value
  • Do NOT infer units based on domain knowledge

Calculated Data Traceability

When creating calculated values, always link them to their source data using data-source-aggregate-document:

{
    "calculated-data-name": "DIN",
    "calculated-result": {"value": 5.8, "unit": "(unitless)"},
    "calculated-data-identifier": "TEST_ID_147",
    "data-source-aggregate-document": {
        "data-source-document": [{
            "data-source-identifier": "TEST_ID_145",
            "data-source-feature": "sample"
        }]
    }
}

This declares: "DIN 5.8 was calculated from the sample at TEST_ID_145."

Why this matters:

  • Audits: Prove a value came from specific raw data
  • Debugging: Trace unexpected results back to their source
  • Reprocessing: Know which inputs to re-analyze if algorithms change

Assign unique IDs to:

  • Measurements, peaks, regions, and calculated values
  • Use a consistent naming pattern (e.g., INSTRUMENT_TYPE_TEST_ID_N)

This enables bidirectional traversal: trace from calculated → raw, or raw → all derived values.


Nested Document Structure (Critical)

A common mistake is "flattening" fields directly onto measurement documents when they should be wrapped in nested structures. This breaks schema compliance and loses semantic context.

Why Nesting Matters

ASM uses nested documents for semantic grouping:

Document Purpose Contains
sample document What was measured Sample ID, locations, plate identifiers
device control aggregate document How instrument operated Settings, parameters, techniques
custom information document Vendor-specific fields Non-standard fields that don't map to ASM

Sample Document Fields

These fields MUST be inside sample document, not flattened on measurement:

// ❌ WRONG - Fields flattened on measurement
{
  "measurement identifier": "TEST_001",
  "sample identifier": "Sample_A",
  "location identifier": "A1",
  "absorbance": {"value": 0.5, "unit": "(unitless)"}
}

// ✅ CORRECT - Fields nested in sample document
{
  "measurement identifier": "TEST_001",
  "sample document": {
    "sample identifier": "Sample_A",
    "location identifier": "A1",
    "well plate identifier": "96WP001"
  },
  "absorbance": {"value": 0.5, "unit": "(unitless)"}
}

Fields belonging in sample document:

  • sample identifier - Sample ID/name
  • written name - Descriptive sample name
  • batch identifier - Batch/lot number
  • sample role type - Standard, blank, control, unknown
  • location identifier - Well position (A1, B3, etc.)
  • well plate identifier - Plate barcode
  • description - Sample description

Device Control Document Fields

Instrument settings MUST be inside device control aggregate document:

// ❌ WRONG - Device settings flattened
{
  "measurement identifier": "TEST_001",
  "device identifier": "Pod1",
  "technique": "Custom",
  "volume": {"value": 26, "unit": "μL"}
}

// ✅ CORRECT - Settings nested in device control
{
  "measurement identifier": "TEST_001",
  "device control aggregate document": {
    "device control document": [{
      "device type": "liquid handler",
      "device identifier": "Pod1"
    }]
  },
  "aspiration volume": {"value": 26, "unit": "μL"}
}

Fields belonging in device control:

  • device type - Type of device
  • device identifier - Device ID
  • detector wavelength setting - Wavelength for detection
  • compartment temperature - Temperature setting
  • sample volume setting - Volume setting
  • flow rate - Flow rate setting

Custom Information Document

Vendor-specific fields that don't map to standard ASM terms go in custom information document:

"device control document": [{
  "device type": "liquid handler",
  "custom information document": {
    "probe": "2",
    "pod": "Pod1",
    "source labware name": "Inducer",
    "destination labware name": "GRP1"
  }
}]

Liquid Handler: Transfer Pairing

For liquid handlers, a measurement represents a complete transfer (aspirate + dispense), not separate operations:

// ❌ WRONG - Separate records for aspirate and dispense
[
  {"measurement identifier": "OP_001", "transfer type": "Aspirate", "volume": {"value": 26, "unit": "μL"}},
  {"measurement identifier": "OP_002", "transfer type": "Dispense", "volume": {"value": 26, "unit": "μL"}}
]

// ✅ CORRECT - Single record with source and destination
{
  "measurement identifier": "TRANSFER_001",
  "sample document": {
    "source well location identifier": "1",
    "destination well location identifier": "2",
    "source well plate identifier": "96WP001",
    "destination well plate identifier": "96WP002"
  },
  "aspiration volume": {"value": 26, "unit": "μL"},
  "transfer volume": {"value": 26, "unit": "μL"}
}

Pairing logic:

  1. Match aspirate and dispense operations by probe number
  2. Create one measurement per matched pair
  3. Use source_* fields for aspirate location
  4. Use destination_* fields for dispense location
  5. Include both aspiration volume and transfer volume

Quick Reference: Nesting Decision

Is this field about...

THE SAMPLE BEING MEASURED?
├── Sample ID, name, batch → sample document
├── Well position → sample document.location identifier
├── Plate barcode → sample document.well plate identifier
└── Source/destination locations → sample document (with prefixes)

INSTRUMENT SETTINGS?
├── Standard settings → device control aggregate document
└── Vendor-specific → custom information document

A MEASUREMENT VALUE?
└── Direct on measurement document (e.g., absorbance, volume)

TRANSFER OPERATION TYPE?
└── DON'T use "transfer type" - pair into single measurement
    with source/destination fields instead

Validation

Use validate_asm.py to check for nesting issues:

python scripts/validate_asm.py output.json --reference known_good.json

The validator checks for:

  • Fields incorrectly flattened on measurements
  • Missing sample document wrapper
  • Missing device control aggregate document wrapper
  • Missing custom information document for vendor fields
  • Liquid handler: separate transfer types instead of paired records

Sources

Source: SKILL.md on GitHub

1 warning16d5 checks · Risk SAFE
  • Gen Agent Trust Hub16d

    This skill provides tools for converting laboratory instrument data into standardized Allotrope Simple Model (ASM) formats. It uses established scientific libraries and includes features for data validation and provenance tracking, which align with security best practices for data integrity.

  • Socket16d

    No alerts

  • Snyk16d

    Risk: LOW · No issues

  • Runlayer7mo

    11/11 files flagged

  • ZeroLeaks5mo

    Score: 93/100 · 2 sections analyzed

Signed by skilld at 7c35640. This ties the file your Agent reads to that commit on GitHub. It does not review the instructions.

Last checked against GitHub last week.

Activeupdated 8 months ago

README badge

README badge for anthropics/knowledge-work-plugins/instrument-data-to-allotrope